HIOKI 3158 AC WITHSTANDING VOLTAGE TESTER
Jumlah Pesanan:
Cara PembayaranTransfer Bank (T/T)
Kemas & Pengiriman1 unit
Negara AsalJepang

Keterangan

Guarantees electrical and electronic equipment Safety

This electrical safety test instrument series is designed for insulation resistance and voltage withstand testing of electrical devices and components according to various safety standards. A multitude of automation and labor-saving features are provided to ensure effective testing for a wide variety of requirements and test conditions. Select the most appropriate model for your applications.
- Safety standards compliance testing
- Stores up to 20 sets of test conditions
- Voltage comparator functions
- Zero-V Switching
- Electrical Discharge function
- Automated testing support
- Interlock Protect function
- Selectable remote controls

Basic specifications :

Test function : Voltage endurance test
Testing voltage : 0 to 2.5 kV / 0 to 5.0 kV AC, dual-range configuration
Voltage testing method : Zero-toggle switch
Transformer capacity : 500 VA (maximum 30 minutes)
Voltage adjustment : Manually adjusted slidac
Waveform : Mains waveform
Frequency : Mains synchronous
Measurement items : Voltage: 0 to 5 kV AC (Average value rectified, effective value display), Current: 0.01 to 120 mA AC (Average value rectified, effective value display)
Judgment function : Contents: UPPER-FAIL / PASS / LOWER-FAIL (Digital setting window comparator method)
Timer section : Setting range: 0.5 to 999 seconds
Interfaces : EXT I/O, EXT SW, RS-232C
Display : Fluorescent tube display (digital), Analog meter
Monitor function : Output voltage, detection current
Power supply : 120 V AC, 50/60 Hz (3158-01), 220 V AC, 50/60 Hz (3158-03), 230 V AC, 50/60 Hz (3158-04),
240 V AC, 50/60 Hz (3158-05)
Dimensions, mass : 320 mm(12.60 in)W × 155 mm(6.10 in)H × 263 mm(10.35 in)D
3158-01: 16 kg (564.4 oz), 3158-03/-04/-05: 18 kg (634.9 oz)
Accessories : H.V. TEST LEAD 9615 (high voltage side and return, 1 each), Power cord (1), spare fuse (1)